How to Work with Slope and Intercept in FOSS IQX™

Working with slope and intercept adjustments in FOSS IQX™ allows you to maintain precise and accurate measurement results across your instruments. Follow these steps closely:

app pre.iqx .net device management slope intercept settings 1
Slope/intercept settings menu is found in the Slope/Intercept module

Step 1: Create a Sample Set

  • Access Instrument Data:
  • Navigate to Instrument Data in FOSS IQX™.
  • Select your instrument and filter samples according to your requirements (product type, dates, reference availability).
  • Create Sample Set:
  • Select samples you want to include.
  • Click Add to Sample Set.
  • Choose Create New Sample Set, give it a descriptive name, and set it as static for slope-intercept adjustments.

Step 2: Manage Your Sample Set

  • Review and Edit Sample Sets:
  • Go to Sample Set module.
  • Ensure your set is labeled as static and has the slope-intercept purpose checked.
  • Manage Reference Values:
  • Click Manage Reference Values.
  • Manually input or import reference values through provided Excel templates.
Sample Sets module menus and data.
Sample Sets module menus and data.

Step 3: Slope and Intercept Adjustment

  • Navigate to Slope Intercept Module:
  • Go to the Slope Intercept module in FOSS IQX™.
  • Set Adjustment Scope:
  • Define your scope (instrument-parameter or product-parameter).
  • Be cautious when changing these settings as they may reset previous adjustments.
  • Inspect Current Slope/Intercept Values:
  • Select your instrument, product, and parameter.
  • Review current slope and intercept values.
Main menu of Slope Intercept with 4 areas to modify.
Main menu of Slope Intercept with 4 areas to modify.

Step 4: Conduct the Validation and Adjustment

  • Select Sample Set:
  • Choose the static sample set you prepared earlier.
  • Perform Adjustment Calculation:
  • Validate samples to get adjustment recommendations.
  • Inspect plots (reference plots, residual plots, XY plots) for validation.
  • Check for outliers; exclude samples if necessary.
  • Implement Recommended Adjustments:
  • Accept recommendations provided by FOSS IQX™.
  • Confirm adjustments to update slope and intercept values.
Summary of the Device within the Sample Sets module menu.
Summary of the Device within the Sample Sets module menu.

Step 5: Publish the Update Session

  • Create an Update Session:
  • Go to Instrument Settings.
  • Select the relevant instrument or group and click Publish Update Session.
  • Validate and Confirm:
  • Name your update session clearly (e.g., “Slope adjustment – Protein June 2025”).
  • Validate the session and publish it.
  • Instrument Synchronization:
  • Allow your instruments to synchronize automatically or manually synchronize immediately on the instrument.

Best Practices

  • Regularly perform slope and intercept validations.
  • Clearly document adjustments with descriptive names.
  • Always convert dynamic sample sets to static before adjustments to maintain consistency and accuracy.

Following these steps in FOSS IQX™ ensures accurate, validated measurements, aiding effective quality management across your instruments.